Q1 gw_product_detail_bar.png gw_product_detail_bar.png If I accidentally press the Delete all functions in the memory setting and delete all data. Is there any method to save the data been deleted?

If you press the Delete all, the information will be completely removed, can not be restore again, when using this feature need to pay more attention.

Q2 gw_product_detail_bar.png gw_product_detail_bar.png Does the AFG-3000 interface support Plug and Play (Plug and Play) function?

The default interface provided by the AFG-3000 is RS-232 interface. This interface doesn’t provide plug and play function. But if the interface is set to USB Port, the system will active the Plug and Play function, then only through combining with downloaded driver(website download area), the device can be installed. As for the GPIB interface, supported software is needed to detect the status of this device.

Q3 gw_product_detail_bar.png gw_product_detail_bar.png How to comprehend the comparison of range accuracy between 2MHz Function Generator GFG-8215A and 5MHz Function Generator GFG-8255A.

The range accuracy (5% +1Hz) designed for GFG-8215A is due to the measurement accuracy set by scroll knob instead of the readings shown on display. As to the empty statement of range accuracy for GFG 8255A ,we usually refer to its “Time Base specification” of frequency counter portion to get the accurate tuning.

Q4 gw_product_detail_bar.png gw_product_detail_bar.png In terms of 15MHz Programmable Function Generator GFG-3015, the GFG-3015's output is relating to 50Ω. And how to make right setting if we would like to measure the amplitude by the oscilloscope with high Z input?

The customer could use an adaptor (High Z: 1MΩ~ 50Ω) to connect GFG-3015. GWInstek have provided this useful adaptor ADP-101 which is BNC (J/F) 75 Ω ~ BNC (P/M) 50 Ω. With this useful adaptor ADP-101, they will save the confused issue in measuring double Amplitude.

Q5 gw_product_detail_bar.png gw_product_detail_bar.png Why is the duty cycle specification specified at the frequency lower than the maximum main output frequency for all function generators?

If the duty cycle is set to significant ratio, say 99:1 at 4MHz ,the 1% of frequency of 4MHz will be much higher than 4MHz,thus it is over the bandwidth the FG specified. Therefore, the lower frequency limit of duty cycle specification is indicated.

Q6 gw_product_detail_bar.png gw_product_detail_bar.png How to properly perform the sweep function in GFG-8200 series function generators?

GFG-8200 series function generators can perform the frequency sweep ratio up to 100:1.To properly perform the 100:1 sweep, the steps are as follows. Step1:Select a frequency range, set the frequency to be the maximum frequency of the sweep (stop frequency). Step2:Adjust the SPAN to maximum position. Step3:Perform the sweep function.

Q7 gw_product_detail_bar.png gw_product_detail_bar.png What is the relationship between the sampling frequency and output frequency of a arbitrary waveform?

The relationship is as follows.Where fo is output frequency Fs is sampling frequency np is number of data point. fo=fs/np

Q8 gw_product_detail_bar.png gw_product_detail_bar.png Why the output signal displayed on oscilloscope is 1Vpp, but when feeding the signal to DUT(Device Under Test),the amplitude is only 0.5 Vpp ?

That’s because the input impedance of oscilloscope is 1MΩ, but function generator’s output impedance is 50Ω.
If the DUT load is also 50Ω and it is connected to a function generator, the oscilloscope will measure half of the amplitude. For function generators like AFG-2225, MFG-2000 and AFG -3000 series, etc. the displayed amplitude is based on 50Ω load.
For example, if the displayed is 1Vpp that means the voltage on the connected 50Ω load. If we disconnect the load and use oscilloscope to measure the voltage directly, 2Vpp will be measured by the oscilloscope.
The setting output voltage value on the AFG is defined as the value when the load used is 50ohm. When the load used is not 50ohm, a different output voltage will produce.

Q9 gw_product_detail_bar.png gw_product_detail_bar.png Why has more measurement applications that use DDS function generator to be better than traditional function generator?

DDS FG is better than traditional FG in most of the applications. For example: the narrow pulse, ARB....etc.Because of the frequency stability and low distortion, the DDS is more adaptive as a signal source. Only in the few cases like higher frequency triangle wave and external FM function, the traditional function generators have better functionality.

Q10 gw_product_detail_bar.png gw_product_detail_bar.png In all of type DDS function generators, why the triangle waveform maximun output frequency were always lower than sine waveform and square waveform?

In order to have straight charge and discharge curve of triangle wave, the number needed of data point is more than sine and square waves. In the case of less data point of triangle wave, the loss pass filter will make it look like sine wave.

Q11 gw_product_detail_bar.png gw_product_detail_bar.png What the standard interface of AFG-3000?

USB Host/Device, RS-232, GPIB

Q12 gw_product_detail_bar.png gw_product_detail_bar.png How much maximum output capacity the AFG-3000 does provide Sync Output function, the is how much?

The maximum output capacity is ≥4 TTL load