Q1 | The meter does not match the real value. |
The following scenarios are possible:The following scenarios are possible:
• Make sure the rear panel OVP setting is in the “Normal” position. If the OVP switch is in the “SET” position, the meter might show the OVP setting, not the output value. For OVP details, please refer user manual page22.
• If you are using Channel 1 or Channel 3, make sure the meter switch on the right side is in the correct position. Channel 1 and 3 share the same meter.
Q2 | The OVP value/ OCP status has not been changed. |
After editing OVP value and OCP status, press the (Vset/Iset/Enter) key to confirm. Otherwise the setting does not change.
Q3 | The actual output is lower than the setting. |
The cable loss might be causing the load voltage lower than the setting. Upgrade the output cable to a better capacity or use the remote sense to compensate for the voltage drop. Please refer user manual.
Q4 | The load wire does not fit into the terminal. |
Make sure you are NOT connecting the load wire to the remote sensing terminal. Please Refer our user manual page30 for connection details.
Q5 | What is the key lock function of the GPD-3303S / GPD-3303D? |
The LOCK key to lock the front panel key operation. When the LOCK key function enable that the LED turns on. To unlock, press the LOCK key for 2~3 seconds then the key LED will be turn off.
Q6 | The CH3 overload indicator turned on is this error? |
When the output Current level exceeds 3A, the overload indicator turns red and CH3 operation mode switches from Constant Voltage to Constant Current. Note:"overload"in this case does not mean an abnormal operation please remove your load then overload indicator would be turns off if you have higher voltage and current requirement you can shift to CH1 or CH2 for operating.