2025 EDM

 

Measurement Applications of the DAQ-9600

 

Creating a Dedicated Multi-Point Current Measurement Module

 

 

The DAQ-9600 data acquisition system offers seven modules for implementing multi-channel measurement.

Among them, the DAQ-901 and DAQ-909 modules each provide two channels capable of directly measuring current.

 

 

Module description Type

Speed

(ch/sec)

Max

volts

Max

amps

Bandwidth

Thermal

offset

Comment
               

DAQ-900

20 ch Multiplexer

2-wire solid-state

(4-wire selectable)

450 120 V   10 MHz < 4 μV Built-in cold junction reference

DAQ-901

20 ch Multiplexer + 2 ch current

 2-wire armature

(4-wire selectable)

80  300 V 1 A 10 MHz < 4 μV

Bulit-in cold junction reference

2 additional current channels (22 total)

DAQ-903

40 ch Single-Ended Mux

1-wire armature

(common low)

80 300 V   10 MHz < 1 μV No four-wire measurements

 DAQ-904

4 x 8 Matrix

2-wire armature   300 V   10 MHz < 1 μV  

DAQ-907

Multifunction Module

16 bit of digital

input and output 

  42 V       Open drain
100 kHz totllizer input    42 V   100 kHz   Input threshold selectable
Two 18-bit analog outputs    ±12 V ±24 mA      Max 40 mA total output per frame

DAQ-908

20 ch Actuator / General Purpose Switch

SPDT / from C    300 V   10 MHz < 4 μV  

 DAQ-909

8 ch HV Multiplexer + 2 ch current

2-wire armature

(4-wire selectable) 

60 

DC 600 V

AC 400 V

2 A  10 MHz < 4 μV

High voltage / high current channels

2 additional current channels (10 total)

 

 

When multiple current measurement points are required, a common approach is to connect a shunt in series and measure the voltage across them (converting current to voltage).

 

However, measurement discrepancies often occur due to variations in the tightness of the connections between the shunt and the measurement channels. The DAQ-909 (8-channel high-voltage + 2-channel high-current multiplexer) includes designated solder pads on the module for mounting shunt resistors.

 

By soldering the shunts directly onto the pads, measurement errors caused by loose screw connections can be eliminated. As a result, this module effectively becomes a dedicated current measurement module. 

 

Note: Recommended shunt values

1 Ω : For milliampere (mA) current measurements

0.1 Ω: For ampere (A) current measurements (up to 2 A)

 

When using the DAQ-Data Logger software, set the function to DCV, and use the MX+B conversion feature to convert the measured voltage into current (A).

 

(Settings) 

 

 

 

(Results) 

  